Device programs:
1. Rapid sterilization program - single use of the absorber; temp. 134 ° C; pressure 2.1 bar; sterilization program time 3 min; drying time 3 min
. 2. Sterilization program for the set of tools - triple use of the absorber; temp. 134 ° C; pressure 2.1 bar; sterilization program time 5 min; drying time 10 min
. 3. Sterilization program for cotton or selected tools - triple use of the absorber; temp. 134 ° C; pressure 2.1 bar; sterilization program time 15 min; drying time 8 min
. 4. Sterilization program of rubber or selected tools - single use of the absorber; temp. 121 ° C; pressure 2.1 bar; sterilization program time 15 min; drying time 4 min
The device includes:
- device
- 3 trays
- handle for removing the trays
- hose for draining and pouring water
- power cord
Technical data:
Basic:Rated voltage: AC 220V, 230V, 50Hz
Rated power: 1500VA
Fuse: 220V: F12AL
Working temperature: 5 ~ 40 ℃
Atmospheric pressure range: 70 kPa ~ 106 kPa;
Plate weight: 4000 N / m2
Noise: <70db
Maximum capacity of one tray: 1000g
Maximum use time of the load test: 90 minutes
Maximum heat radiation energy under the conditions of 20 ℃ ~ 26 ℃: <2000J
Chamber:
Material: stainless steel (for medical use)
Max.working pressure: 2.5bar Min.working
pressure: -0.9bar Max.temperature
: 145
komory Chamber volume: 18L (Φ245 × 350mm)
Loading size: 18L (Φ232 × 325)
Max load weight: 18L (3.07kg / cm2) 23L
Working pressure / temperature: 1.10 ~ 1.30 bar / 121 ℃ ~ 122 ℃; 2.10 ~ 2.30 bar / 134 ℃ ~ 135 ℃
Water volume per cycle: 0.16 l ~ 0.18 l
Sterilizer steam safety valve:
Safety release pressure: 2.45 bar
Max. Working temperature: 160 ℃
Water tank:
Main water tank capacity: 18 L (3.5 L)